首页 | 本学科首页   官方微博 | 高级检索  
相似文献
 共查询到1条相似文献,搜索用时 0 毫秒
1.
This study examined the validity of the Actical accelerometer step count and energy expenditure (EE) functions in healthy young adults. Forty-three participants participated in study 1. Actical step counts were compared to actual steps taken during a 200 m walk around an indoor track at self-selected pace and during treadmill walking at different speeds (0.894, 1.56 and 2.01 m · s–1) for 5 min. The Actical was also compared to three pedometers. For study 2, 15 participants from study 1 walked on a treadmill at their predetermined self-selected pace for 15 min. Actical EE was compared to EE measured by indirect calorimetry. One-way analysis of variance and t-tests were used to examine differences. There were no statistical difference between Actical steps and actual steps in self-selected pace walking and during treadmill walking at moderate and fast speeds. During treadmill walking at slow speed, the Actical step counts significantly under predicted actual steps taken. For study 2, there was no statistical difference between measured EE and Actical-recorded EE. The Actical provides valid estimates of step counts at self-selected pace and walking at constant speeds of 1.56 and 2.01 m · s–1. The Actical underestimates EE of walking at constants speeds ≥1.38 m · s–1.  相似文献   

设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号