Performance evaluation of the high sensitive troponin I assay on the Atellica IM analyser
Authors:Antonio Buo Soto  Katell Peoc&#x;h  Tommaso Fasano  Jorge Diaz-Garzon  Bernardino Gonzlez de la Presa  Valerie Chicha-Cattoir  Simone Canovi  Maria Sanz de Pedro  Nayra Rico  Tiphaine Robert  Efrem Bonelli  Pilar Fernndez Calle  Aurea Mira  Guillaume Lefevre  Luigi Vecchia  Jose Luis Bedini
Abstract:IntroductionThe Fourth Universal Definition of Myocardial Infarction Global Taskforce recommends the use of high sensitive troponin (hs-Tn) assays in the diagnosis of acute myocardial infarction. We evaluated the analytical performance of the Atellica IM High-sensitivity Troponin I Assay (hs-TnI) (Siemens Healthcare Diagnostics Inc., Tarrytown, USA) and compared its performance to other hs-TnI assays (Siemens Advia Centaur, Dimension Vista, Dimension EXL, and Abbott Architect (Wiesbaden, Germany)) at one or more sites across Europe.Materials and methodsPrecision, detection limit, linearity, method comparison, and interference studies were performed according to Clinical and Laboratory Standards Institute protocols. Values in 40 healthy individuals were compared to the manufacturer’s cut-offs. Sample turnaround time (TAT) was examined.ResultsImprecision repeatability CVs were 1.1–4.7% and within-lab imprecision were 1.8–7.6% (10.0–25,000 ng/L). The limit of blank (LoB), detection (LoD), and quantitation (LoQ) aligned with the manufacturer’s values of 0.5 ng/L, 1.6 ng/L, and 2.5 ng/L, respectively. Passing-Bablok regression demonstrated good correlations between Atellica IM analyser with other systems; some minor deviations were observed. All results in healthy volunteers fell below the 99th percentile URL, and greater than 50% of each sex demonstrated values above the LoD. No interference was observed for biotin (≤ 1500 µg/L), but a slight bias at 5.0 g/L haemoglobin and 50 ng/L Tn was observed. TAT from was fast (mean time = 10.9 minutes) and reproducible (6%CV).ConclusionsReal-world analytical and TAT performance of the hs-TnI assay on the Atellica IM analyser make this assay fit for routine use in clinical laboratories.
Keywords:analytical techniques and equipment  high-sensitive troponin I  immunoassay  performance evaluation
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号