首页 | 本学科首页   官方微博 | 高级检索  
     检索      


An Evaluation of the Point-of-Care Analyzer,i-Smart 30, for Measurement of Electrolytes
Authors:Subhash C Arya  Nirmala Agarwal  Beena Michael
Institution:1. Sant Parmanand Hospital, 18 Alipore Road, Delhi, 110054, India
Abstract:Analytical functioning of a point-of-care analyzer, i-Smart 30 (i-sens: Seoul, South Korea), for electrolyte quantification was investigated at Sant Parmanand Hospital, a tertiary-care hospital in Delhi, India. Samples that were received for electrolyte assay were assayed, double-blinded for their Na and K level using the arterial blood gas analyzer, the ABL 555 (Radiometer, Copenhagen) and the i-Smart 30 electrolyte analyzer. There was satisfactory correlation between the results obtained with the two analyzers with an encouraging bias, standard deviation and the 95 % limits of agreement between the data generated for Na and K levels. The performance of the i-Smart 30 would be satisfactory during the point-of-care measurements of Na and K levels in emergency rooms and clinical laboratories with inadequate infrastructure only if its day-to-day performance was monitored to ensure reliability of the generated reports.
Keywords:Electrolytes  Point-of-care assays  Field studies  Emergency investigations  Laboratory infrastructure
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏