首页 | 本学科首页   官方微博 | 高级检索  
     检索      


Validierung von neun Bewegungssensoren bei Kindern und jungen Erwachsenen
Authors:Dr Carsten Müller  Corinna Winter  Marie-Sophie Mogwitz  Dieter Rosenbaum
Institution:1. Funktionsbereich Bewegungsanalytik, Institut für Experimentelle Muskuloskelettale Medizin (IEMM), Universit?tsklinikum Münster, Domagkstr. 3, 48149, Münster, Deutschland
Abstract:The purpose of this study was to determine the accuracy of one accelerometer and eight pedometers in schoolchildren and young adults under controlled experimental conditions and under free-living conditions over a 24-h period. The most accurate results under controlled conditions were obtained with a piezoelectric pedometer. For this device absolute deviations (median) ranged between 0.2 and 1.0% for children and between 0.2 and 1.4% for young adults. Absolute deviations for the accelerometer were 0.8–3.1% for children and 0.6–3.2% for adults. Motion sensors were more accurate with increasing walking speed. For several pedometers the absolute deviation exceeded 50% in tests with schoolchildren. These differing results illustrate that step counts from different motion sensors cannot be compared directly. Therefore, the validity and reliability of activity monitors should be examined carefully before being used in scientific research.
Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏